CPR Savers offers two styles of conserving devices.
Our O2 pneumatic conserving regulator Part# 130- 0500 & 130- 0505 is 100% conserver or 100% regulator depending on the dial position located on top of the unit. The flow settings are easily dialed-in from 1 LPM up to 6 LPM to meet your patient's needs. The dual lumen cannula technology is utilized to cover the vast majority (over 95%) of the patient population making it trouble-free for your respiratory therapists and technicians. Since the conserver is pneumatic, no batteries or electronics are involved to confuse the patient. In fact, our Respond O2 conserver is lighter and smaller than the vast majority of electronic and pneumatic units on the market.
Hours of Use |
|
|
|
|
|
|
(Conserve mode, based on 20 breaths/min) |
|
|
|
|
|
|
|
|
|
|
Cylinder Type |
Capacity |
1 LPM |
2 LPM |
3 LPM |
4 LPM |
6 LPM |
M6, ML6 |
164 L |
8.2 |
4.1 |
2.7 |
2.1 |
1.3 |
M9, C |
248 L |
12.5 |
6.2 |
4.2 |
3.1 |
2.1 |
D |
425 |
21.3 |
10.4 |
6.9 |
5.2 |
3.5 |
E |
684 |
34.2 |
17.1 |
11.4 |
8.5 |
5.7 |
Oxygen Use |
|
|
(ml/ breath @ LPM setting |
|
|
|
|
LPM Setting = ml/ Breath |
LPM Setting = ml Breath |
LPM Setting = ml Breath |
1.0 = 28 ml |
3.0 = 61 ml |
5.0 = 95 ml |
1.5 = 37 ml |
3.5 = 70 ml |
5.5 = 103 ml |
2.0 = 46 ml |
4.0 = 78 ml |
6.0 = 111 ml |
2.5 = 53 ml |
4.5 = 86 ml |
|
- Light weight aluminum outer sleeve with brass inner core 13.8 oz.
- May be used on carbon composite cylinders to 3000 PSIG
- Test to CGA/ASTM standards including ignition testing
- Flow range and settings: 1-6 LPM; 0,1,1.5,2,2.5,3,3.5,4,4.5,5,5.5,6 LPM
- High visibility 1 1/2" diameter gauge
- 5 year warrenty -- Made in the USA
|